3B G3 Series User Manual page 42

Table of Contents

Advertisement

Luna G3 BPAP 25A (G3 B25A) User Manual
Problem
Water in mask
Nasal, sinus, or
ear pain
Discomfort due
to
inability
to
adapt
to
the
treatment
pressure
Obstructive
sleep
apnea
symptoms recur
The
device
is
too noisy
Air
delivered
from the device
is
abnormally
hot
Possible Cause
When the humidifier is used,
the humidified air tends to
condense
in
the
tubing and mask if the
room temperature is low.
Sinus
or
middle
inflammation
treatment
pressure
prescription, and cannot
treat sleep apnea if the
treatment pressure is set
too low.
There are many causes of
recurring
symptoms
including weight change,
medication or alcohol, and
poor mask fit.
The
tubing
is
connected properly.
The air inlet of the device
may be partially blocked,
leading
to
insufficient
airflow into the device.
39 / 48
Solution(s)
Turn the humidity setting down
or raise the room temperature.
Place the tubing under the quilt
or use the tubing cover.
cold
Hang the tubing loosely, the
lowest part of the tubing should
be lower than the patient's
head.
ear
Contact
immediately.
Patients may experience a
is
a
variance in time to acclimate
to their therapy.
If the problem persists, contact
your physician or home care
company for assistance.
Contact
your
assistance.
not
Reconnect the tubing properly.
Replace the air filter (see 18.7
Replacing the Air Filter / PM2.5
Filter) and clean the air inlet.
Place the device in an area
where air flows freely, and
make sure the device is at least
20 centimeters away from the
wall, curtain, or other things.
MID-97/v1.0
your
physician
physician
for