Exiting The Patient Menu; Options Of The Patient Menu And Corresponding Descriptions - 3B G3 Series User Manual

Table of Contents

Advertisement

Luna G3 BPAP 25A (G3 B25A) User Manual

16.1.7 Exiting the Patient Menu

The users can press the Home
16.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Humidifier
Preheat
Reslex
Heated
Tubing
Ramp Time
Delay
Date
to return to the Main Interface shown in Fig. 16-1.
Range
There are five humidity levels available. As the
Off,Auto,
numbers
accordingly. "Off" means the humidifier is
1 ~ 5
turned off.
Set humidifier to preheat by adjusting this
On / Off
option. This feature automatically turns off after
30 minutes.
This feature enables the device to automatically
reduce the treatment pressure when the patient
exhales, so as to make the user more comfortable.
Off, 1 ~ 3
The higher the number is, the more pressure the
device reduces. "Off" means this feature is
disabled.
There are five heat levels available. As the
number increases, the heat rises accordingly.
Off, 1 ~ 5
"Off" means the heat is turned off. Note:
Heated Tubing is displayed in the patient menu
only when it is connected.
In order to increase comfort and help the
patient fall asleep easily, the pressure can
increase graduallywhen the Ramp feature is
enabled. The ramp time during which the initial
Auto, 0 -
pressure rises to the prescribed treatment
pressure can be adjusted. As you turn the Knob
Max Ramp
to the nearest point, the numbering
increases or decreases by five minutes. The
screen displays a real-time countdown of the
remaining ramp time in seconds.
When the humidifier is on, this feature allows the
airflow to continue for about 15 minutes at a
low pressure (about 2 cmH
the
treatment. This will blow off the vapor left in the
On / Off
water chamber to avoid any damage to the
device. When this feature is set to "Off," which
means it is disabled, the airfolw stops delivering
air instantly after you press the Start / Stop
2000-01-01
Set date by adjusting this option.
2099-12-31
30 / 48
Description
increase,
the
Start
/
Stop
.
MID-97/v1.0
humidity
rises
O) after you press
2
to
discontinue