Luna G3 BPAP 25A (G3 B25A) User Manual
(3) Returning the Transfer Box: As shown in Fig. 18-3.
CAUTION!
• It is recommended to clean the transfer box once a week.

18.4 Cleaning the Enclosure

Wipe the surface of the device with a soft, slightly damp cloth.
CAUTIONS!
• The device should only be used after the enclosure is dry, so that no moisture
enters the device.
• It is recommended to clean the enclosure once a week.

18.5 Cleaning the Tubing

(1) Remove the tubing from the device and mask before cleaning.
(2) Clean the tubing in warm water and mild soap, and then rinse it in clean water
thoroughly.
(3) After cleaning, air-dry the tubing in a cool, well-ventilated area, and avoid direct
sunlight. It takes approximately 30 minutes to completely air-dry the tubing. Ensure
the tubing is completely dry before re-use.
CAUTION!
• It is recommended to clean the tubing once a week.

18.6 Replacing the Air Filter / PM2.5 Filter

(1) Open the air filter cap to remove the air filter. Attach the air filter to the filter cap,
as shown in Fig. 18-4.
(2) Install the filter cap containing the air filter to the device, as shown in Fig. 18-5.
Transfer Box
Fig. 18-3
Air Filter
Filter Cap
Fig. 18-4
35 / 48
MID-97/v1.0