Hameg HMC8012 Programmer's Manual page 10

Scpi commands
Hide thumbs Also See for HMC8012:
Table of Contents

Advertisement

SCPI Commands HMC8012
Numeric values
Numeric values can be entered in any form, i.e. with sign, decimal point and exponent. Values exceeding the resolution
of the instrument are rounded up or down. The mantissa may comprise up to 255 characters, the exponent must lie
inside the value range -32000 to 32000. The exponent is introduced by an „E" or „e". Entry of the exponent alone is
not allowed. In the case of physical quantities, the unit can be entered. Allowed unit prefixes are MA (mega) / MOHM,
K (kilo), M (milli) and U (micro). If the unit is missing, the basic unit is used.
S C P I P r o g r a m m e r s M a n u a l
CALC:NULL:OFF 10mV
Example:
Special numeric values
S C P I P r o g r a m m e r s M a n u a l
S C P I P r o g r a m m e r s M a n u a l
The texts listed below are interpreted as special numeric values. In the case of a query, the numeric value is provided.
– MIN/MAX
MINimum and MAXimum denote the minimum and maximum value.
Example:
Setting command:
Query:
Queries for special numeric values
The numeric values associated to MAXimum/MINimum can be queried by adding the corresponding mnemo-
nics to the command. They must be entered following the quotation mark.
Example:
Returns the maximum numeric value as a result.
Boolean parameters
Boolean parameters represent two states. The „ON" state (logically true) is represented by „ON" or a numeric value
1. The „OFF" state (logically untrue) is represented by „OFF" or the numeric value 0. The numeric values are provided
as the response for a query.
:ACQuire
Example:
:CHAN1OFFSet
Setting command:
Query:
Response:
Text parameters
Text parameters observe the syntactic rules for mnemonics, i.e. they can be entered using a short or long form. Like
any parameter, they have to be separated from the header by a white space. In the case of a query, the short form of
the text is provided.
Example:
Setting command:
Query:
Response:
HAMEG Instruments GmbH • Industriestr. 6 • D-63533 Mainhausen • Deutschland
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: info@hameg.com
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: info@hameg.com
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: info@hameg.com
Geschäftsführer: Dipl.-Ing. Holger Asmussen, Dipl.-Ing. Andre Vander Stichelen • AG Offenbach am Main HRB 41200
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
=
CALC:NULL:OFF 10E-3
CALCulate:LIMit:LOWer MAXimum
CALC:LIM:LOW MAX?, Response: 1E3
CALC:LIM:LOW? MAXimum
CALCulate[:STATe] ON
CALC:STAT?
1
:DISPlay
:HCOPy
HCOPy:FORMat BMP
*ESR?
HCOPy:FORMat?
BMP
H M O S e r i e s
H M S S e r i e s
H M S S e r i e s
Firmware Version: 01.020 and later
Firmware Version: 01.020 and later
Firmware Version: 03.00 and later
:MEASure
:LOGic:SIZE
*OPC?
HMO352x, HMO2524
HMO72x ... HMO202x
:POD1
:SYSTem
*SRE?
Subject to change without notice
Subject to change without notice
Subject to change without notice
English
English
English
10
1
1
1

Advertisement

Chapters

Table of Contents
loading

Table of Contents