Usage Guidance and restrictions:
• Maximum altitude 2,000m.
• Temperature range between 4°C and 65°C.
• Maximum relative humidity 80% for temperatures up to 31
C decreasing
O
linearly to 50%relative humidity at 40
C.
O
• Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: "Normally only non-conductive
pollution occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected".
Care and Maintenance:-
Cleaning Horizontal Units
Units are best cleaned using warm water and a mild detergent. Water at
temperatures above 60
C can cause damage to the unit and
0
components.
The tank should be thoroughly rinsed with warm water or distilled water to
prevent build up of salts but care should be taken not to damage the
enclosed electrode and vigorous cleaning is not necessary or advised.
Air drying is preferably before use.
The units should only be cleaned with the following:-
Warm water with a mild concentration of soap or other mild detergent.
Compatible detergents include dishwashing liquid, Hexane and Aliphatic
hydrocarbons
The units should not be left to in detergents for more than 30 minutes.
The units should never come into contact with the following cleaning
agents, these will cause irreversible and accumulative damage:-