Provide adequate airflow
Do not block the air intake at the front and side of the instrument or the exhaust at
the rear. Install the instrument on a location that allows sufficient space for air cir-
culation at the air intake and exhaust. Recommended spacing to non-heat pro-
ducing surface is at least 2.5 inches (63.5 mm) from the ventilation holes.
Applicable local or national safety regulations and rules for the prevention of acci-
dents must be observed in all work performed.
The instrument is designed for use in the following sectors: Industrial, residential,
business and commercial areas and small businesses.
The instrument is designed for indoor use only. Before each measurement, you
need to verify at a known source if the instrument functions properly.
To disconnect from the mains, the low-heat device socket on the back panel
has to be unplugged.
See
Table 3-1
for the general data on the instrument specification. For more infor-
mation, see the instrument product brochure.
Table 3-1: General data on instrument specification
Mains nominal voltage
Power consumption
Mains fuses
(fuse size: 5 mm x 20 mm)
Temperature
Humidity
Display
Rack mount capability
Dimensions
Getting Started 5601.8902.02 ─ 06
AC
Maximum input power
100 V / 115 V AC
230 V AC
Operating temperature range
Storage temperature range
Non-condensing
W x H x D
Putting into Operation
Intended Operation
100 V / 115 V / 230 V (±10 %)
50 Hz / 60 Hz
230 W
IEC 60127 T5.0H250V
IEC 60127 T2.5H250V
5 °C to + 40 °C
- 20 °C to + 70 °C
5 % to 80 %
3.5 " (QVGA)
R&S HZN96 rack adapter 2U
(P/N: 3638.7813.02)
222 mm x 96 mm x 446 mm
(8.74 in x 3.78 in x 17.56 in)