B. Braun Aquaboss CCS Instructions For Use Manual page 72

Table of Contents

Advertisement

10.
Safety markings
10.1
Existing markings and lettering checked for
damaged and replaced if necessary
11.
Inspection of the earthing equipment
connections
11.1
Earth terminal fitted tightly to rail
11.2
Clamps on earth terminal tightened
11.3
Earthing bolts for door tightened
11.4
Earthing bolts for mounting plate tightened
11.5
Earthing bolts for housing tightened
12.
Junction boxes
12.1
Tighten all clamps on junction box FU1
12.2
Tighten all clamps on junction box FU2
12.3
Tighten all clamps on junction box FU3
13.
Repeat test for electric medical devices in accordance
with DIN EN 62353 (IEC 62353)
Test equipment number:
ATTENTION
Test executed in accordance with DIN EN 62353 (IEC 62353)
13.1
Visual inspection:
Mains connection cable, complete system:
13.2
Protective conductor resistance:
13.3
Measurement between mains connection cable and housing
13.4
Measurement between mains connection cable and mounting plate
13.5
Measurement between mains connection cable and door/front plate
13.6
Measurement between mains connection cable and
ground connection of concentrate pipe
13.7
Insulation resistance:
Gauge all conductive parts that can be touched
13.8
Leakage current:
Equivalent leakage current:
Functional test
Measuring protocol available
Executed
14.
Handover
14.1
Start dialysis / standby operating mode
14.2
Attach test label
Next STK due at the latest:
STK + certificate: The CCS has been checked for correct functionality. The CCS is safe to operate under the conditions of proper use.
.................................................................................................
Service technician, block letters
The system was taken over in perfect working order
.................................................................................................
System operator, block letters
Rev. 2.8 Date 2019-02-04
Software version 1.00
Replaced
Carried out /
Last replacement
OK
month / year
Replaced
Carried out /
Last replacement
OK
month / year
Replaced
Carried out /
Last replacement
OK
month / year
Carried out /
OK
(MM/YY)
.................................................................................................
City / date, signature
.................................................................................................
City / date, signature
Medical product log Aquaboss
Section 2 – Operating Instructions
Values / data / remarks
Values / data / remarks
Values / data / remarks
Values / data / remarks
OK
n.OK
Measured value Limit value
R
< 0,300 
SL
R
< 0,300 
SL
R
< 0,300 
SL
R
< 0,300 
SL
R
> 2,0 M
ISO
I
< 1,0 mA
EGA
OK
n.OK
OK
n.OK
OK
n.OK
CCS
®
CCS
R
V
X
X
R
V
X
X
X
X
X
X
X
X
X
X
R
V
X
X
X
X
X
X
W
V
X
X
X
X
X
X
R
V
X
X
X
Page 7-5