9.

Faults / Causes / Correction

9.1

Fault messages

CCS-YX.X
Error 0.0-1
CPU
Menu
9.1.1

Types of faults

There are 3 types of fault messages:
1. ALARM
A deviation from normal operation has been established. The effects can impair system operation.
The control system is not switched off automatically.
Indicated by:
Reduced operation is continued
(see table on next page, "effects" column)
Fault message display
Collective alarm (relay)
Reset via menu or self-acknowledgement (depending on fault)
The alarm can reset itself automatically when operating states change.
2. ERROR
A fault has been established. The effects could lead to system damage. The system is switched off
automatically.
Indicated by:
System is switched off
Fault message display
Collective alarm (relay)
Reset by switching the system off and on or via menu
(depending on the fault)
NOTE
3. MALFUNCTION
Malfunction of an individual line. The supply line affected is switched off. The other lines continue
to operate normally.
Indicated by:
Fault message display
Collective alarm (relay)
Reset via menu or self-acknowledgement (depending on fault)
Rev. 2.8 Date 2019-02-04
Software version 1.00
When a fault is determined in the system, the following appears in
the display:
hh:mm
xx stands as a placeholder for the different fault codes.
An error message overwrites an alarm message. The first error message is retained even
if further "errors" are established.
Faults / Causes / Correction
Section 1 – Operating Instructions
CCS
Page 9-1