B. Braun Aquaboss CCS Instructions For Use Manual page 58

Table of Contents

Advertisement

Check ring for continuity
Line 1.1
Line 1.2
Line 2.1
Line 2.2
Line 3.1
Line 3.2
Protective conductor resistor /
electr. safety DIN EN 61010-1
test equipment number
Measurement betw. connection cable
and housing
Measurement betw. connection cable and
mounting plate
Measurement betw. connection cable and
doors/front panel
Measurement between connection cable and
earthing connection for concentrate line
Insulation test
Leakage current
Housing leakage current normal condition
(NORMAL CONDITION)
Housing leakage current first fault
(SIGNAL FAULT)
Lauer Membran Wassertechnik GmbH
Rev. 2.8 Date 2019-02-04
Software version 1.00
C
Carried out /
Left
OK
not OK
OK
not OK
OK
not OK
OK
not OK
OK
not OK
OK
not OK
OK
Carried out /
OK
Industriegebiet Speichermatt
OMMISSIONING LOG
Right
not OK
OK
not OK
OK
not OK
OK
not OK
OK
not OK
OK
not OK
OK
Values/data/remarks
NR
Measured value
R
SL
R
SL
R
SL
R
SL
R
ISO
I
GA.NC
I
GA.SF
79599 Wittlingen
Germany
Section 2 – Operating Instructions
Values/data/
remarks
R
V
Limit value
<0.1
X
<0.1
X
<0.1
X
<0.1
X
>2M
X
<0.1mA
X
<0.5mA
X
Initial commissioning
CCS
Page 4-7

Advertisement

Table of Contents
loading

Table of Contents