Lifespan of the Medical Device

This medical device was designed for a service life of 10 years, excluding wear parts in normal working
conditions. This duration may change depending on the frequency of use.
CHECK AT LEAST ONCE A YEAR THE GENERAL CONDITION OF THE DEVICE.
The wear parts are :
• The upholstery
• The knobs and screws
Promotal recommends replacing the wear parts after a maximum of 5 years of use.
For any intervention, please contact your local authorized dealer, indicating the serial number of the device.

Fuse replacement

1.
Remove all power to unit by unplugging unit's power cord.
2.
Using a fl at tipped screwdriver, gently release the plastic spring clip (A) on both sides of fuse drawer (1)
then pull fuse drawer from fuse housing (2).
3.
Pull both fuses (3) out of fuse drawer (1) and inspect. Check the fuses for any indication that they have
blown ; i.e. burnt look, fuse cord melted through, etc. Discard fuses (3) if blown and replace them.
EQUIPMENT ALERT
Use fuses of the same voltage rating, amperage rating, and type. Failure to
do so could result in damage to the equipment.
4.
If necessary, obtain new fuse(s) (3). The replacement fuse(s) must be a 250 VAC, 6.3 amp,
IEC 127 rated, 5 x 20 mm, Type T "Slo-Blo".
5.
Insert fuse(s) (3) into fuse drawer (1).
6.
Insert fuse drawer (1) into fuse housing (2) until fuse drawer snaps into place (both side of use drawer
are locked into fuse housing).
7.
Plug in power cord to equipment. If fuse blows again, call your Promotal distributor.