1.

Introduction

1.A. Maxwell
16 MDx Purification Procedure
®
The Maxwell
16 MDx Instrument
®
The methods use sample lysis and binding to paramagnetic particles as the primary separation principle. It has
two modes for different elution volumes. The Standard Elution Volume (SEV) mode allows elution in a volume of
up to 400µl. The Low Elution Volume (LEV) mode allows elution in as little as 25µl. Up to 16 samples can be
prepared in a single run.
The automated steps performed by the Maxwell
Sample lysis in the presence of a chaotropic agent, detergent and/or alcohol
Binding of nucleic acids or proteins to paramagnetic particles
Washing of the target molecules bound to the particles away from other cellular components
Elution of the product
The instrument contains a touch screen for navigating, programming and running the instrument. It has a UV
lamp to aid with decontamination. The Maxwell
Track Software, has the ability to record and provide run data. It has a USB port that can be used to attach a bar
code reader, allowing sample and reagent information to be entered using bar codes. The Maxwell
Instrument can report the data gathered for each run, and the report can be sent to a serial printer or to a
computer.
The user selects the protocol to be run, collects the selected bar code information (optional) and places the
samples into the reagent cartridges. The cartridges are placed into the machine and the door shut to start the run,
which automatically performs the protocol.
The temperature of the samples is regulated by a heating system that is controlled by the protocol.

Maxwell

®
Sample Track is a communications program that allows downloading and printing of the run data
(see Technical Manual #TM314).
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
provides automated nucleic acid purification methods for multiple analytes.