2 - 4
Cabling, basic unit
Fig. 6 Remove cable cover
•
Remove the cable cover (Fig. 6).
•
Remove the base plate as well (Fig. 7).
remove
unscrew
Fig. 8 Cable support, beam entry
•
Mark the cables with a pen (Fig. 8) so that you are able to set the exact cable length later
on.
•
Take off the cable support at the beam entry (Fig. 8).
•
Take off the tape (Fig. 8).
•
Remove the upper cable support (Fig. 9).
•
Move the lifting column into the top position.
SIREMOBIL Iso-C 3D
Actions up to serial number 1408
mark
SPR2-230.814.01
Rev. 05
Fig. 7 Existing base plate
remove
Fig. 9 Upper cable support
Page 4 of 14
08.04
CS PS 24
2
Siemens AG
Medical Solutions