Installation
Fig. 21

Cables for basic unit

Insert the cables in the energy chain. Make sure that cables do not intersect.
Fasten both cables for the cable module using cable clamps (1/Fig. 21).
Close the energy chain.
Secure the last link on both ends of the energy chain using cable ties or insulation tape to
avoid accidental opening.
Attach the ferrite core to the cable with the plug M2. X1 (1 turn around the ferrite core).
Connect the plug.
t
Attach the connection cable for the inverter. Tighten the nuts with a torque of 4.8 Nm.
Fasten the cable using a cable clamp. Make sure that the shielding makes good ground
contact.
Fasten the protective ground wires.
Attach the Plexiglas cover (4/Fig. 21).
Reconnect the BNC plug for the coaxial cable.
Insert the previously ejected contacts of the plug D30 X3 in the connector housing, as
previously noted.
Run the cables with their plugs to the corresponding plug connections / PC boards ac-
cording to labeling. Fix the cables using cable ties.
Tie the cables close to the lifting column using cable ties in such a way that they do not
rub against any moving parts or are crushed when moving the lifting column.
Fix all cables to the cable holder (1/Fig. 21) using cable ties.
Reattach all shielded cables using the provided cable clamps. Make sure that the shield-
ing makes good contact (2/Fig. 21) (1/Fig. 22).
Siemens AG
Medical Solutions
Fig. 22
SPR2-240.841.01
Rev. 01
05.03
Page 9 of 10
CS PS 24