Fig. 27
When drilling, ensure that the drill is placed vertically against the plate.
Due to the step length of the drill, a small 26 mm diameter bridge remains on the inner
edge of the hole. Remove this bridge with a half-round file. The diameter drilled should
be 28 mm across the entire plate thickness. Use the file to remove the burr as well.
Also deburr the two 3.5 mm holes. This can be done with a drill having a diameter of
approx. 8 mm to 10 mm. Place the drill manually against the 3.5 mm holes from the
inside and brake the burr.
Remove all drill shavings outside and inside with a vacuum cleaner.
Varnish the 28 mm hole with a varnish applicator and allow it to dry.
Position the connector with holding plate (1/Fig. 27) from the inside.
The connector is positioned centrally to the 28 mm hole, the lid opens downward.
Fasten the connector with holding plate using the screws and lock washers supplied
(1/Fig. 28).
Siemens AG
Medical Solutions
2
3
4
SPR2-230.814.01
Rev. 05
Fig. 28