4 - 2
Fig. 5 Cutting the threads
Move the C-arm into the 0º position (Fig. 1).
Apply marks at 20 mm intervals to determine the lateral limit for the cam (Fig. 2).
Position the cam. Ensure that the cam is flush with the top and mark the holes with the
cams laterally at the 20 mm mark (Fig. 3).
Punch-mark the holes.
Ensure that shavings or dust do not drop into the C-arm.
Use a 3.2 mm bit for drilling. Continue until you have have drilled through the wall.
Ensure that you are not drilling beyond the wall (Fig. 4).
Counterbore the holes.
Cut the threads using an M4 thread tap (e.g. use a tap wrench) (Fig. 5).
Attach the cam using 2 countersunk screws M 4 x 8 (Fig. 6).
Repeat these steps on the other side of the C-arm.
SIREMOBIL Iso-C 3D
Kapitelüberschrift hier eintragen
Fig. 6 Attaching the cam