6 - 6
3. Part of activities independent of serial number
TFT Displays
NOTICE
Dismantling existing monitors
•
Both monitors should be removed as follows.
Markierung für Tastaturschublade
Fig. 16 Monitor cover
•
Remove the monitor covers (Fig. 16).
•
Remove the shielding plates (Fig. 17).
Markierung für Tastaturschublade
Fig. 18 Connection housing
•
Remove the connection housing (only for SIMOMED N monitor 11 02 628, Fig. 18).
•
Unplug all cables from the monitor to be replaced.
•
Remove the two screws on each monitor attachment (Fig. 19).
•
Lift both monitors down.
•
Reattach all the parts removed from the monitor.
SIREMOBIL Iso-C 3D
If a SIMOMED HM monitor was ordered and delivered, disregard
this section. Continue with the section "PC support".
SPR2-230.814.01
Rev. 05
Fig. 17 Shielding plate
Fig. 19 Monitor attachment
Page 6 of 42