2.7. Measuring ranges

The METPOINT
FLM can measure flow velocities up to 185.0 m/s and is preconfigured for an inside pipe diameter of 53.1 mm. At
®
the analogue output of 4 ... 20 mA, this corresponds to:
Rated
Inner Ø
diameter
Inch
mm
1/4"
6.0
10.0
15.0
1/2"
16.1
3/4"
21.7
1"
25.0
26.0
27.3
28.5
30.0
1 1/4"
32.8
36.0
36.3
1 1/2"
39.3
40.0
41.9
43.1
45.8
2"
50.0
51.2
53.1
54.5
57.5
60.0
64.2
2 1/2"
65.0
70.3
71.1
76.1
* According to DIN 1945/ISO 1217 (20 °C, 1000 mbar), with compressed air.
** Adjustment for DIN 1343: 0 °C, 1013.25 mbar
METPOINT
SF53
®
Volume flow (full scale value in Nm³/h)
Air *
Air **
9.4
8.7
29.8
27.4
77.7
71.4
91.0
83.7
177.8
163.5
243.9
224.3
265.2
243.9
294.7
271.0
323.3
297.3
361.1
332.0
436.7
401.6
531.5
488.7
541.1
497.6
639.8
588.4
663.7
610.3
728.4
669.8
777.3
714.8
882.2
811.2
1059.2
974.1
1112.1
1022.6
1197.6
1101.3
1263.1
1161.6
1491.6
1371.7
1544.1
1420.0
1774.3
1631.7
1821.0
1674.6
2137.9
1966.0
2186.8
2011.0
2511.2
2309.3
Installation and operating manual
Max.
N
**
m/s
2
8.7
185.0
27.4
185.0
71.4
185.0
83.7
185.0
163.5
185.0
224.3
185.0
243.9
185.0
271.0
185.0
297.3
185.0
332.0
185.0
401.6
185.0
488.7
185.0
497.6
185.0
588.4
185.0
610.3
185.0
669.8
185.0
714.8
185.0
811.2
185.0
974.1
185.0
1022.6
185.0
1101.3
185.0
1161.6
185.0
1371.7
185.0
1420.0
185.0
1631.7
185.0
1674.6
185.0
1966.0
185.0
2011.0
185.0
2309.3
185.0
15