7. Spare parts and accessories

The accessories available for the METPOINT
Designation
Power supply unit with A plug-type
connector
 4032115
High-pressure safety device
 4025892

8. Maintenance and servicing

Regularly check the sensor head for dirt and clean as necessary. Dirt, dust or oil deposits on the sensor element cause incorrect
measurements.
We recommend checking the sensor element at least once a year. If the compressed air is heavily soiled, use a shorter inspection
interval.

9. Sensor head cleaning

To clean the sensor head, immerse it in warm water with a small amount of washing-up liquid. Do not clean the sensor with a cloth,
sponge, brush or other implement, as any mechanical shock can destroy the sensor. If dirt deposits cannot be removed, return the
sensor to the manufacturer for inspection and maintenance.

10. Recalibration

Unless otherwise agreed with the customer, we recommend having the device calibrated every 12 months. For calibration, send
the METPOINT
FLM to BEKO TECHNOLOGIES GmbH.
®

11. LED display

A calibration reminder LED is mounted on the top of the housing of the METPOINT
recalibration becomes due and is indicated by the LED flashing. The flashing LED has no effect on measurement data. The
measurement signal is output as before.
The reminder time interval can be adjusted to customer requirements at the manufacturer's works.
METPOINT
SF53
®
FLM are listed in the table below.
®
Installation and operating manual
Illustration
®
FLM. After 15 months have elapsed,