Basic Setting Commands - YOKOGAWA XL100 User Manual

Portable data station communication function
Hide thumbs Also See for XL100:
Table of Contents

Advertisement

5.6 Setting Commands (Control)/5.7 Basic Setting Commands
GROUP key:
MARK key:
MARK left key:
MARK right key:
SET key:
START/STOP key:
Up arrow key:
Down arrow key:
Right arrow key:
Left arrow key:
SELECT key:
Example
Press the DISPLAY key.
KE DISP
Description Operates in the same fashion as the key
operation on the XL100. For consecutive key
operations, transmit the commands in the same
order as the key operation on the XL100.
BZ
Sets the activation of the setting
commands.
BZ p1<terminator>
Syntax
p1 Activation method of setting commands
Activate after the setting command is
executed (default): 1
Activate when the settings are saved using
the XE command after executing the setting
command: 0
BZ?
Query
Example
Activate the setting commands when the settings
are saved using the XE command as with basic
setting commands.
BZ 0
Description • The default setting (the response to BZ? is BZ
1) is always in effect at power-on.
• This command is activated immediately.
5-32
GROUP
5.7
MARK
MARKLEFT
MARKRIGHT
• In order to activate the settings that are changed
SET
STARTSTOP
UP
DOWN
RIGHT
LEFT
SELECT
• Basic setting commands cannot be used to change
XA
XI

Basic Setting Commands

using the basic setting commands, the settings
must be saved using the XE command.
The settings that are returned in response to a
query after issuing a basic setting command are the
new settings even if the settings have not been
saved with the XE command. However, the new
settings only take effect after saving the settings
using the XE command.
If you discard the settings with the XE command or
start the logging operation before saving the
settings with the XE command, the settings that are
returned in response to a query are reset to those
before the change.
settings while logging is in progress.
Sets the alarm
XA
Syntax
p1,p2,p3,p4,p5,p6,p7,p8,p9<terminator>
p1 Reserved
p2 Output number to be set to AND
For all ORs:
Independent setting:
Range setting: I0x-I0y
Multiple settings (delimit using a period):
p3 Reserved
p4 Output hold/non-hold
Hold:
Non-hold:
p5 Hold/Not hold the alarm status display
Hold:
Non-hold:
p6 Reserved
p7 Reserved
p8Alarm hysteresis ON/OFF: ON or OFF
p9Alarm buzzer ON/OFF:
XA?
Query
Example
Set output numbers I01 to I04 to AND, hold the
output, hold the alarm status display, turn the
alarm hysteresis ON, and turn the alarm buzzer
ON.
XA ,I01-I12,HOLD,HOLD,,,ON,ON
Description Parameters p1, p3, p6, and p7 are not used.
Sets the filter.
XI p1<terminator>
Syntax
NONE
I01, I02, I03, or I04
(I0x,I0y:I01 to I04)
I0w.I0y.I0z
(I0w,I0y,I0z:I01 to
I04)
HOLD
NONHOLD
HOLD
NONHOLD
ON, OFF
IM XL100C-E