7.2

Error Messages Related to Operation

External Storage Media Operation Errors
Code
Message
200
Operation aborted because an error was found in media.
201
Not enough free space on media.
202
Media is read-only.
210
Media has not been inserted.
211
Media is damaged or not formatted.
212
Format error.
213
The file is read-only.
214
There is no file or directory.
215
Exceeded the allowable number of files.
216
The file or directory name is incorrect.
217
Unknown file type.
218
Directory exists. Delete the directory or change
directory name.
219
Invalid file or directory operation.
220
The file is already in use. Try again later.
230
There is no setting file.
231
Abnormal setting exists in file.
Review Display Errors
Code
Message
232
There is no available data.
233
The specified review data do not exist.
234
The specified channel is not assigned to the display group. –
IM XL100C-E
This section gives a list of error codes and messages of the XL100 and their corrective
action. Error responses to communication commands are output in English.
Description/Corrective Action
Check the external storage medium.
Replace the external storage medium.
Set the external storage medium to enable writing.
Insert an external storage medium.
Replace the external storage medium or format it.
Format the external storage medium again.
Access another file or make the file write-enable.
Replace the external storage medium.
Delete unneeded files.
Specify another file.
Tried to delete multiple directory levels.
Wait until the file is accessible.
Specify another file.
Specify another file.
Description/Corrective Action
Displayed when showing the review display.
Specify another file.
Displayed when showing the review display.