Usb Communication Specifications And Setup Procedure - YOKOGAWA XL100 User Manual

Portable data station communication function
Hide thumbs Also See for XL100:
Table of Contents

Advertisement

3.3
USB Communication Specifications and Setup
Procedure
Connection Procedure of the USB Communication Interface
Setting the USB Interface
3-8
When using the USB for serial communication, the XL100 is connected as a device to a
host calculater such as a PC.
The only communication protocol that can be used is normal protocol (Yokogawa Meters
& Instruments proprietary protocol).
Client End
Item
Number of ports
Electrical and
mechanical specifications
Connector
Power supply
PC system supported
Connect a USB cable to the Mini-B connector on the XL100 and a USB hub or a type A
connector of a PC on the host calculater end.
Set the following item.
Setting the USB ID
Set the USB ID number of the XL100 within the following range.
00 to 31
You can connect multiple devices to a host controller on the USB. If the XL100 is
connected to multiple devices in a single USB system, the USB ID number is used by the
host controller to identify each device. Therefore, unique ID numbers must be assigned
to the XL100s within a single system.
Note
Do not change the USB ID number while using the USB.
Procedure
1.
Press SETTING to enter Setting Mode.
Specifications
1
Conforms to USB Rev.1.1
5-pin Mini-B receptacle
Self-powered
A PC running Windows 2000 or Windows XP that is equipped
with a USB port as standard (a separate device driver is
required for the connection with a PC)