Output Commands (Setting/Measurement/Data Output) - YOKOGAWA XL100 User Manual

Portable data station communication function
Hide thumbs Also See for XL100:
Table of Contents

Advertisement

5.8 Output Commands (Control)/5.9 Output Commands (Setting/Measurement/Data Output)
Description Parameter p4, the terminal block serial number,
is undefined if the terminal block unit is not
connected.
IT
Outputs terminal block status
(query only).
IT p1<terminator>
Syntax
p1 Terminal block status
When the 8-ch terminal block is connected
0
When the 16-ch terminal block is connected
1
When the terminal block is not connected
3
Query
IT?
IM XL100C-E
5.9
Output Commands (Setting/
Measurement/Data Output)
FC
Outputs the screen image data
Syntax
FC p1<terminator>
p1 Screen image data output GET
Example
Output the screen image data from the XL100.
FC GET
Description Captures the current displayed screen on the
XL100 and outputs the data in BMP format.
FE
Outputs the setting data
Syntax
FE p1,p2,p3,p4,p5,p6,p7,p8,p9,p10,
p11<terminator>
p1 Output data type
Setting data of setting commands:
Decimal place and unit information:
Setting data of basic setting commands: 2
Reserved:
Reserved:
p2 First channel number
For 8 ch input:
For 16 ch input:
p3 Last channel number
For 8 ch input:
For 16 ch input:
p4 First pulse measurement channel number:
PL01
p5 Last pulse measurement channel number:
PL01
p6 First logic input channel number
DI01 to DI02
p7 Last logic input channel number
DI01 to DI02
p8 First calculation channel number
CA01 to CA32
p9 Last calculation channel number
CA01 to CA32
p10 First communication channel number
CO01 to CO32
p11 Last communication channel number
CO01 to CO32
Example
Get the decimal place and unit for channels 01 to
12 and pulse channel.
FE 1,01,12,PL01,PL01
Description • Set the first channel number and last channel
number parameters so that the last channel
number is greater than or equal to the first
channel number.
• Parameters p2, p3, p4, p5, p6, p7, p8, p9,
p10, and p11 are valid when p1 is set to 0, 1,
or 2.
0
1
3
4
01 to 08,
01 to 16
<Numeric value> 00(OFF)
01 to 08,
01 to 16
<Numeric value> 00(OFF)
5-43
1
2
3
4
5
6
7
App
Index

Advertisement

Table of Contents
loading

This manual is also suitable for:

Im xl100c-e

Table of Contents