Channel bandwidth setup
If for example a signal is applied to CH1 which is a pulse signal that contains high
frequency oscillation.
Press CH1 to select Channel1.
Press F2 to set the BANDWIDTH LIMIT
OFF so it is set up as full bandwidth.
The signal being measured can now pass
through even if it contains high frequency
quantities.
Press F2 to set BANDWIDTH LIMIT ON so
that frequency quantities higher than 20MHz in the signal being measured will be
limited.
Vertical Volts/Div adjustment setup
The VOLTS/DIV range of the vertical deflection factor can be adjusted either in
coarse or fine tune mode.
In COARSE TUNE the VOLTS/DIV range is
2mV/div~5V/div. Tuning is in steps of 1-2-5.
In FINE TUNE mode the deflection factor
can be adjusted in smaller steps allowing
continuous adjustment within the range
2mV/div~5V/div without interruption.
Probe rate setup
To match the probe attenuation factor setup, it is necessary to set up the probe
attenuation factor in the channel operation
menu accordingly.
For example when the probe attenuation
factor is 10:1, set the probe attenuation
factor at 10X in the menu. This principle
applies to other values to ensure the voltage
reading is correct.
Waveform inversion setup
The displayed signal is inverted 180
degrees with respect to the ground level.
7