INSTALLATION INSTRUCTIONS
2.
Connect power to the MINILOADER.
The wires in this mains lead are coloured in accordance with the following code:
As the colours of the wires in the mains lead of this appliance may not correspond with the coloured markings
identifying the terminals in your plug proceed as follows:
The wire coloured green and yellow must be connected to the terminal on the plug marked with the letter E
or by the earth symbol
The wire coloured brown must be connected to the terminal marked with the letter L or coloured red.
The wire coloured blue must be connected to the terminal marked with the letter N or coloured black.
3.
Insert an INTERLOCK OVER-RIDE KEY into the TOP COVER INTERLOCK SWITCH. Turn the
MINILOADER MAIN SWITCH on. The DISPLAY should read "PLEASE WAIT-RESET OPERATING"
and the ORANGE LIGHT should be on. The CAM MOTOR should run until the, "HOME" position is
reached. The GREEN LAMP should come on and the DISPLAY should read, "MINILOADER READY
FOR FORMAT 18*24".
If the MINILOADER does not operate correctly, consult the SERVICE MANUAL.
4.
Press the OPERATOR BUTTON, "MAGAZINES TO RELOAD POSITION"
Open the FRONT DOOR and install the 18 x 24 cm MAGAZINE, loaded with TEST FILM into the upper
MAGAZINE TRAY and the 24 x 30 cm MAGAZINE, loaded with TEST FILM, into the lower MAGAZINE
TRAY. Close the FRONT DOOR. The MINILOADER is now ready for testing.
Check the PUSH RODS etc. for tightness.
5.
Clean the SUCKERS by rubbing them with grade 200 EMERY CLOTH, or a similar material, and dust them
6.
with NATURAL TALCUM POWDER. If the surface of the SUCKERS is subsequently touched, this
procedure must be repeated.
Vacuum out the MINILOADER 2 PLUS to remove any dust.
7.
Fit the STEP-BY-STEP SWITCH , run a cycle with each size of film and check the SUCKER VACUUM
8.
is between 150 & 200 millibar. Adjust if necessary. Check the COMPRESSOR reaches a minimum pressure
of 2.0 bar. Correct if necessary.
Run several cycles with each available size of TEST FILM, checking for correct TILT operation, good
9.
FILM separation in the SUPPLY MAGAZINE, correct insertion of FILMS into the CASSETTES and
correct delivery of FILM into the CHUTE (or RECEIVING MAGAZINE, depending on version of
MINILOADER). Make any adjustments necessary.
January 1995
IMPORTANT
GREEN AND YELLOW - EARTH
BLUE
BROWN
or coloured green or green-and-yellow.
Page II-25
- NEUTRAL
- LIVE
II 3420
© Kodak Ltd.