INSTALLATION INSTRUCTIONS
10.
Align the MINILOADER and PROCESSOR horizontally as shown in figure 16, fit the other two M4
LARGE DIAMETER WASHERS and the M4 NUTS. Do not tighten the NUTS yet. It is very important
to have the alignment correct, especially for an RA PROCESSOR because the CHUTE EXTENSION
can interfere with the LED's inside the FILM ENTRY.
Figure 16
HORIZONTAL
ALIGNMENT
11.
By adjusting the LEVELLING SCREWS on the MINILOADER, (and the PROCESSOR if necessary),
set the height of the MINILOADER. The vertical alignment is correct when a FILM hits the lower
ENTRY ROLLER 1.5 mm below the nip of the ROLLERS. See figure 17.
Figure 17
VERTICAL
ALIGNMENT
12.
When both horizontal and vertical alignments are correct, tighten the M4 NUTS.
13.
Go to, "INSTALLING THE HUMIDIFIER" on page II-22 of this publication to complete the
installation..