(ii) Changing from LEV to Standard Configuration

1.
Ensure that the power is turned off and the instrument is unplugged. This will release the motors so that the
heads can be gently moved to allow easier access to the hardware assemblies.
2.
Gently and slowly, with constant pressure on both the right and left sides, push the black LEV Magnetic Rod
Assembly down to the lowest position on the slides. Do not push the LEV Magnetic Rod Assembly too fast as
this could result in damage to the instrument's electronics.
3.
Unscrew the three thumbscrews on top of the magnetic
rod assembly.
4.
Gently lift up the LEV Magnetic Rod
Assembly to remove it.
5.
Pull down on the LEV Plunger Bar Adaptor to
release the magnets that hold it in place.
Hint: Pull down at an angle so that the
back of the magnet is released first.
Remove the black LEV Plunger Bar Adaptor.
6.
Insert the blue SEV Magnetic Rod Assembly.
7.
Secure the Magnetic Rod Assembly with the three thumbscrews, and hand-tighten.
8.
Proceed to Section III.D, and complete the configuration by updating the Firmware to the
Standard Elution Volume settings.
Failure to change the instrument firmware operational mode from the low elution
!
volume to the standard elution volume settings will cause damage to the instrument.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 6/08
Thumbscrew
LEV Magnetic Rod Assembly
LEV Plunger Bar Adaptor
SEV Magnetic Rod Assembly
Part# TM274
Page 15