Air tubing:

General
The patient is an intended operator.
Operator position
The device is designed to be operated within arm's length. An operator should position their line of sight within an angle of
30 degrees from a plane perpendicular to the screen.
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
Heated air tubing temperature cut-out: ≤ 43°C
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the nasal cannula.
• Do not use electrically conductive or antistatic air tubing.
• The temperature and relative humidity settings displayed are not measured values.

Displayed values

Value
Flow
1
Results are expressed as STPD (Standard Temperature and Pressure, Dry). (101.3kPa at an operating temperature of 20°C, dry). Use the
table below to convert the STPD flow setting to BTPS (Body Temperature and Pressure, Saturated) flow.
BTPS Flow (L/min)
Altitude (m)
0
500
1000
1500
2000
2500
Measurement system uncertainties
In accordance with ISO 80601-2-74:2017 the measurement uncertainty of the manufacturer's test equipment for measures
of flow is ± 1.5 L/min or ± 2.7% of reading (whichever is greater)
Guidance and manufacturer's declaration electromagnetic emissions and
immunity
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
The Lumis HFT device has been designed to meet EMC standards. However, should you suspect
that the device performance (eg, pressure or flow) is affected by other equipment, move the device
away from the possible cause of interference.
22
6 months
Accuracy
1
±6 L/min at 0 to 40 L/min positive flow
Target Flow Setting (L/min)
15
20
17
23
18
24
19
26
20
27
22
29
23
31
Length
2 m
25
30
28
34
30
36
32
38
34
41
36
43
38
46
Inner diameter
15 mm
35
40
39
45
42
48
45
51
48
54
51
58
54
62