Installing the probe

Coaxial probe
1.
Make sure the process connection is at least ¾" NPT or a flanged mounting.
2.
Carefully place the probe into the vessel. Align the gasket on flanged installations.
3.
Align the probe process connection with the threaded or flanged mounting on the
vessel.
4.
For threaded connections, tighten the hex nut of the probe process connection. For
flanged connections, tighten flange bolts.
Twin Rod Probe (Rigid)
active
probe
rod
min.
25 mm
(1")
threaded or flanged mounting on the vessel.
5.
For threaded connections, tighten the hex nut of the probe process connection. For
flanged connections, tighten flange bolts.
6.
Probe can be stabilized by attaching the inactive probe rod to vessel.
Page 30
inactive
probe
rod
SITRANS LG200 – INSTRUCTION MANUAL
1.Make sure the process connection is at
least 2" NPT or a flanged mounting.
2.Make sure that there is at least 25 mm (1")
spacing between the active probe rod and
any part of the tank (walls, still-well, pipes,
support beams, mixer blades, etc.). Minimum
stillwell or nozzle diameter for Twin Rod
probe is 80 mm (3").
3.Carefully place the probe into the vessel.
Align the gasket on flanged installations.
4.Align the probe process connection with the
7ML19985KA02