Keysight X Series Installation Note page 7

Agile signal generators
Hide thumbs Also See for X Series:

Advertisement

If no SSD is desired in the instrument:
3. Install the empty drive tray by sliding the tray into the rear panel opening. Secure by
tightening the thumb screw. Torque to 9 in-lbs.
For a customer purchased SSD installation:
4. Refer to
Figure
2. Remove the SSD plate (2) by using the T-10 driver to remove the 4 screws
(1).
Figure 2
Plate Removal
Installation Note N5191-90025
1CC145A - Empty Drive Tray Accessory Installation
7