Advanced configuration
December 2019
9. In Source Device Type / ID / No list, select the appropriate HART device.
10. Choose the appropriate source parameter; Hart PV, Hart SV, Hart TV, or Hart QV.
11. In case there are more HART slaves connected, repeat
12. In TankMaster WinSetup, install and configure the tank that the HART Slave device
13. Verify that the tank parameter receives measurement data from the HART Slave by,
C.10.3
HART slave mapping
In case a Rosemount 2240 Multi-Input Temperature Transmitter is used, some parameters
can not be mapped manually to a HART Slave as described in section
configuration. See
188
Example
Vapor Pressure is selected in this example.
Example
In the example above, a 3051 PT is selected.
was associated with in the Rosemount 2410 tank database (see step
for example, opening the Tank View window in TankMaster WinOpi.
Table C-6
for a list of parameters that can be mapped.
Reference Manual
00809-0100-2410
Step 4
to
Step
7.
Step
3).