Battery Replacement - Covidien ForceTriad Service Manual

Energy platform
Hide thumbs Also See for ForceTriad:
Table of Contents

Advertisement

Battery Replacement

Equipment required:
• Phillips screwdriver
Warning
Electric Shock Hazards To allow stored energy to dissipate after power is disconnected, wait at
least five minutes before replacing parts.
Caution
The system contains electrostatic-sensitive components. When repairing the system, work at a
static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized system. Handle PCBAs by their non-
conductive edges. Use an antistatic container for transport of electrostatic-sensitive components
and PCBAs.
Notice
Calibrate the system after every battery replacement. Calibration values are lost when the battery
is replaced. Refer to Calibrating the ForceTriad Energy Platform on page 6-4 for instructions.
1.
Turn off the system. Disconnect the power cord from the wall receptacle.
2.
Remove the four screws that secure the cover to the chassis. Slide the cover off the
chassis. Set the cover and screws aside for reinstallation.
3.
Locate the battery on the left side of the Controller PCBA.
4.
Grasp the battery and remove it from the socket.
5.
Install the new battery.
6.
Position the cover above the chassis and slide it down. Install the four screws that
secure the cover to the chassis.
7.
Perform a level-6 calibration of the system as described in Calibration Levels on page
6-6.
Battery Location on Controller Board
ForceTriad Energy Platform Service Manual
Battery Replacement
8-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents