Start-Up Defaults For Ventilation Parameters And; Alarm Limits - Dräger Medical NeoFlow Addendum To Operating Instructions

Neonatal patient mode for evitaxl
Table of Contents

Advertisement

Start-up Defaults for
Ventilation Parameters and

Alarm Limits

Start-up default values, i.e. the values for
tidal volume V
and ventilator rate f
T
activated whenever Evita
XL
on, may be defined as a function of
either ideal body weight or of the patient
mode.
Press »
System Setup« function
key.
Touch »Therapy« screen key.
Touch »Mode & Settings« screen key
and enter access code 3032.
Touch »V
, f...« screen key.
T
Change specific values as required by
hospital protocol:
Touch the corresponding screen key.
Turn dial knob to set.
Press dial knob to confirm.
To restore factory default settings:
Touch »Dräger Default« screen key .
Press dial knob to confirm.
is switched
Configuration of Ventilation
Start-up Defaults for Ventilation Parameters and Alarm Limits
29