Start-Up Defaults For Ventilation Parameters - Dräger Medical Evita 2 dura Operating Instructions Manual

Intensive care ventilator
Hide thumbs Also See for Evita 2 dura:
Table of Contents

Advertisement

Patient-specific defaults V
Patient
mode
Tidal volume
V
mL
T
Pediatric
50
Adult
500
* Hospital selected values may be entered into the last columns for reference.

Start-Up Defaults for Ventilation Parameters

– These defaults specify the ventilation parameters and
alarm limits that are not patient-specific and are
activated on starting up the device.
Press »Calib./Config.« menu key.
Press »Ventilation jj« menu key.
Enter code number »3032«.
Select the »Parameter« menu with »Ventilation jj«
menu key.
Display (example):
Turn dial knob to select the desired screen field,
press dial knob to confirm.
Turn dial knob to set value,
press dial knob to confirm.
, f:
T
Factory-set
Ventilator rate
f bpm
29
12
Hospital-specific settings*
Tidal volume
Ventilator rate
V
mL
f bpm
T
......................................
......................................
......................................
......................................
Leak
comp.
Configuration
Ventilation Defaults at Start-Up
99