Anesthesia Ventilator Operation ............5-12 Part Number: 4115139-001 Narkomed Mobile Operator’s Manual Rev: F...
Page 4
Breathing Pressure Monitoring Problem Resolution .........8-6 Appendix: Spare and Replacement Parts Rev: F Part Number: 4115139-001 Narkomed Mobile Operator’s Manual...
Page 5
Recommendations ............. 1-4 Purpose of This Manual ............. 1-4 Symbol Definition ............... 1-5 Warnings and Cautions ............. 1-5 Warnings ..............1-6 Moving the Unit ............1-6 Cautions ............... 1-8 Vaporizer Storage ............1-9 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Draeger Medical, disclaims responsibility for the adequacy of the monitoring package selected for use with the anesthesia system. However, Draeger Medical is available for consultation to discuss monitoring options for different applications. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Datagrip, DrägerService, Narkomed, Narkomed GS, ORM, Quality Service For Life, Vigilance Audit, Vitalert, and Vitalink are registered trademarks of Draeger Medical, Inc. All other products or name brands are trademarks of their respective owners. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
For further information concerning these agreements, contact DrägerService at (800) 543-5047. Purpose of This Manual This manual provides operating instructions for the Narkomed Mobile Anesthesia System. It is intended for use by trained clinical professionals familiar with accepted medical procedures, practices, and terminology used in delivery of anesthesia and patient monitoring.
Warning statements provide important information that, if ignored, could lead directly to personal injury. • Caution statements provide important information that, if ignored, could lead directly to equipment damage and indirectly to personal injury. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Contact DrägerService at (800) 543-5047 for further information. Moving the Prior to moving the Narkomed Mobile unit, the following functions must be Unit performed: 1. Disconnect the power cord from the mains and store properly.
Page 11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Figure 1-1. Narkomed Mobile Showing Handle Locations. ROTATE MONITOR HANDLE INTO PARKED POSITION (SIDE) HANDLE (REAR) ROTATE ABSORBER, ELEVATE SLIGHTLY, AND INSERT INTO BRACKET OPENING FOR PARKED POSITION...
Do not insert any additional components into, or modify, the anesthesia system after any checkout procedure is started. Doing so can invalidate the checkout results and could affect the patient's safety. There is a possible explosion hazard if the Narkomed Mobile is used in the presence of flammable anesthetics. Cautions...
There are two additional foam cushions supplied with the machine and are located in the drawer. Figure 1-2. Vaporizer Storage Locations SU00045 ADDITIONAL FOAM PAD (IF NEEDED) VAPORIZER HOLDERS Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 15
Adjustable Pressure Limiter (APL) Valve ....2-17 Respiratory Volume Monitor Sensor ......2-17 Oxygen Sensor ............2-18 Positive End-Expiratory Pressure (PEEP) Valve ..2-18 Scavenger System ............2-19 Anesthesia Ventilator ............2-20 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 16
Remote Display and Support Arm ......... 2-24 Monitoring System ............2-25 Monitor Screen and Controls ........... 2-25 Alarm System ..............2-26 Alarm Display ..............2-26 Alarm Annunciation ............2-27 Ventilation Alarms ............2-27 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Overview The Narkomed Mobile is a compact, lightweight, continuous flow anesthesia system. All Narkomed Mobile machines are equipped with a monitoring system and pneumatic circuitry for delivering gases and anesthetic vapor for adult and pediatric patients.
Page 18
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS The Narkomed Mobile consists of these major systems: • gas delivery • vaporizer • absorber • scavenger • ventilator • power supply • system interface • monitoring/alarms.
Each connection, valve, gauge, and flowmeter is labeled and color-coded for the appropriate gas, as shown in the table below. GAS SYSTEM COLOR CODING MARKING COLOR Yellow Nitrous Oxide Blue Oxygen Green Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 20
The gauges are located directly below their corresponding flowmeters and flow control valves. Pressure is indicated in psi and kPa. The pressure gauge and flowmeter arrangements are shown in Figure 2-4 on page 2-7. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 21
O cylinders and deplete their contents (if the cylinder valve is left in the open position). Caution: To ensure gas supplies are adequate, pipeline pressure gauges should show steady pressures of 50—55 psi. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 22
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Cylinder Gas The Narkomed Mobile is equipped with a two cylinder hanger yoke Entry assembly. The standard combination is one O cylinder and one N cylinder. A factory installed option for the mounting of two O cylinders is available.
The specific gas is labeled at bottom of the flowtube. Each flowmeter has a float indicator. To determine the flow rate, read the flowmeter scale at the center of the float. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
The oxygen dispensing system has a calibrated bypass flow of Oxygen Flow 150 ±50 mL/min (at 50 psi pipeline pressure) that delivers this flow of oxygen even if the oxygen flow control valve is fully closed. 2-10 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Figure 2-6. Flush Button OP00097 O2 FLUSH FRESH GAS CONTROL OUTLET FRESH GAS LOCKING BAR RELEASE TO LOCK PULL TO INSERT OR REMOVE HOSE FRESH GAS HOSE Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-11...
Flowmeter on the left side of the flowmeter bank (Figure 2-6 on page 2-11). This flowmeter can be used when the machine is turned off. 2-12 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 27
RETURN TO CD-ROM TABLE OF CONTENTS Figure 2-7. Fresh Gas Outlet OP00097 O2 FLUSH FRESH GAS CONTROL OUTLET FRESH GAS LOCKING BAR RELEASE TO LOCK PULL TO INSERT OR REMOVE HOSE FRESH GAS HOSE Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-13...
RETURN TO CD-ROM TABLE OF CONTENTS Anesthetic Vaporizer The Narkomed Mobile is equipped with a Dräger-Vapor 19.3 vaporizer. This vaporizer enriches fresh gas with precisely metered vapor of liquid anesthetic agent. The vapor is inserted into the fresh gas line. It is connected between the fresh gas metering unit and the fresh gas outlet.
The absorber system handles spontaneous, manually assisted, or automatic ventilation. The preferred mode of operation is selected with the manual/ automatic selector valve. The “AUTO” position of the valve enables automatic ventilation. The “BAG” position enables Manual/Spontaneous ventilation. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-15...
A cap is tethered to the absorber for plugging into the inspiratory valve dome when the oxygen sensor and pressure sensor hose assembly are not in place. 2-16 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Valve absorbent. Respiratory The respiratory volume sensor is an ultrasonic flow sensor that monitors Volume tidal volume, respiratory minute volume, and respiratory frequency. Monitor Sensor Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-17...
(lower than –0.5 cmH O) and excessive positive pressure (higher than +10 cmH O) are not possible at the connection point. 2-18 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
A 19 mm hose connects to the fitting below the hose barb for passive exhaust systems. A relief valve guards against excessive pressure build-up if the exhaust hose is restricted. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-19...
32 psi. Electrical power is supplied by the AC power source, or, in event of AC power failure, by the backup battery. A fully charged battery can power the ventilator for at least 90 minutes. 2-20 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 35
(PEEP) within the breathing system. For the Narkomed Mobile, the PEEP is approximately 2 cmH The pressure limit control is used to set the peak inspiratory pressure produced by the ventilator to limit the maximum pressure to the patient.
RETURN TO CD-ROM TABLE OF CONTENTS Power Supply System The Narkomed Mobile has a central power supply for the ventilator, alarm system, and monitoring system. When in use, the Narkomed Mobile must be plugged into an active AC outlet. Do not use “cheater” plugs. The term “cheater”...
The battery test button and indicator located near the system power switch Indicator and are used to determine the backup battery status (Figure 2-13 on page 2-22). Button A detailed testing procedure is included in the checkout procedures in this manual. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-23...
The display can be adjusted up or down to place it in the most convenient position. The support arm below the monitor screen supports an external patient monitor. 2-24 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
RETURN TO CD-ROM TABLE OF CONTENTS Monitoring System The monitoring system integrates the functions of the electronic monitors and organizes information from these monitors on the monitor screen. The Narkomed Mobile monitors: • oxygen concentration • breathing pressure • respiratory volume •...
INSP O2 LOW OXYGEN SUB ATM PRESSURE CAUTION: AC / BATTERY FAIL O2 SUPPLY LOW ADVISORY: SERVICE VENT MON PORT A ERROR TID VOL MIN VOL 0.35 PEAK MEAN PEEP OP87001 2-26 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
When the ventilator is turned back on, the pressure apnea threshold is restored to its previous set value and the apnea alarms revert to activation after 15 seconds (Caution) and 30 seconds (Warning) of apnea. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 2-27...
Page 42
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 43
Oxygen Monitoring ............3-4 Breathing Pressure Monitoring .......... 3-5 Respiratory Volume Monitoring ......... 3-5 Minute Volume ............. 3-5 Tidal Volume ..............3-5 Respiratory Rate ............3-5 Serial Interface ..............3-6 Serial Ports ..............3-6 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
......≥ 8 months at 25 ° C, Sensor service life ≥ . . . 50% relative humidity, 50% O gas mixture (or 5000% hr CO Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Protocols ......... . Vitalink Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 49
APL Valve Verification ..........4-7 Breathing System Leak Test ........4-7 Scavenger System Verification ........4-8 Manual and Automatic Ventilation Systems ....4-9 Monitors ..............4-10 System Flush ............. 4-10 Final Position ............. 4-10 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 50
APL Valve ................. 4-12 Breathing System Leak Test ..........4-13 Scavenger System ............4-14 Manual and Automatic Ventilation Systems ..... 4-15 Monitor ................4-16 System Flush ..............4-16 Final Position ..............4-16 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
RETURN TO CD-ROM TABLE OF CONTENTS Daily Checkout Procedure Before operating the Narkomed Mobile, the following checkout procedure must be performed to make sure the machine is ready for use. This is a recommended procedure. Follow your institution's policies for specific checkout procedures.
Check the flowmeters. Adjust the flow control knob for each gas and Circuit verify the proper operation of the corresponding flowmeters. The float Verification must move freely over the full range of each flowmeter. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Make sure the handwheel is set to 0. Open the oxygen flow control Verification valve to an 8 L/min flow and close all other flow control valves. Sniff the gas coming from the fresh gas common outlet. There should be no noticeable odor. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Wear a surgical mask to inhale and exhale through the breathing system (each limb individually, if possible). Verify the unidirectional flow in each limb and then reconnect the tubing. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Observe the pressure drop at the breathing system pressure gauge. When the pressure is at 50 cmH O, begin counting seconds. The pressure must not drop more than 20 cmH O in 30 seconds. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Turn the PEEP valve control knob fully counterclockwise to its lowest setting. Turn the scavenger needle valve fully counterclockwise. k. Push the O flush button to inflate the scavenger reservoir bag. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Occlude the patient side of the Y-piece. Fill the ventilator bellows by pressing the oxygen flush button. Observe the breathing system pressure gauge as the ventilator cycles. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
BAG f. all flowmeters indicate 0 (or minimum) g. the breathing system is ready to use (the bag is in place and all hoses are connected properly). 4-10 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Make sure the breathing pressure hose assembly is properly connected to the pressure sensor adapter and the system interface panel. g. Make sure the oxygen sensor and respiratory volume sensor are properly installed. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 4-11...
Open the oxygen flow control valve to a flow of 8 L/min. g. Occlude the bag mount opening and watch for a pressure increase on the breathing system pressure gauge. This pressure increase must not exceed 3 cmH 4-12 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Observe the pressure drop at the breathing system pressure gauge. When the pressure is at 50 cmH O, begin counting seconds. Thirty seconds or longer are needed for a pressure drop from 50—30 cmH Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 4-13...
Turn the PEEP valve control knob fully counterclockwise to its lowest setting. Turn the scavenger needle valve fully counterclockwise. k. Push the O flush button to inflate the scavenger reservoir bag. 4-14 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Occlude the patient side of the Y-piece. Fill the ventilator bellows by pressing the oxygen flush button. Observe the breathing system pressure gauge as the ventilator cycles. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 4-15...
0 (or minimum) f. the patient suction level is adequate g. the breathing system is ready to use (bag is in place and all hoses connected are properly). 4-16 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 65
Using the Manual/ Automatic Selector Valve ....5-8 Using the APL Valve ............ 5-9 Using the PEEP Valve ..........5-10 Scavenger Interface Operation ........5-11 Needle Valve Adjustment for Suction Systems ..5-12 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 66
Setting the Respiratory Frequency ........5-15 Setting the Inspiratory/Expiratory (I:E) Phase Time Ratio 5-15 Setting the Inspiratory Flow Rate ........5-15 Setting the Inspiratory Pressure Limit ......5-15 Selecting the Bellows Drive Gas ........5-16 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
RETURN TO CD-ROM TABLE OF CONTENTS Gas Delivery System Operation The Narkomed Mobile is a continuous flow anesthesia system with pneumatic circuitry for mixing and delivering fresh gas and anesthetic agent vapor. The pneumatic system can deliver up to three gases (oxygen, nitrous oxide, and air) and one anesthetic agent simultaneously.
Using the To use the oxygen flush, press the oxygen flush button, located on the front Oxygen of the Narkomed Mobile. This introduces an unmetered flow of pure oxygen Flush into the breathing circuit at a rate of about 55 L/min.
Page 69
Note: Drain the anesthetic agent if the vaporizer will not be used for longer than six months or if the vaporizer will be removed from the anesthesia machine. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Warning: Because the vaporizer funnel filling system does not limit the type of agent poured into the vaporizer, using an agent monitoring device is recommended to verify the agent. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Warning: Do not use the anesthesia machine if: • The pins in the I E valves or valve body(ies) are bent, damaged, or missing, • The valve disks are missing or damaged. • The valve seat is damaged. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 74
PEEP. The actual PEEP in the system is indicated on the absorber pressure gauge at the end of exhalation and can be adjusted from about 2—15 cmH 5-10 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
(if necessary) at six month intervals. Warning: Waste gases vented through the positive pressure relief valve enter the operating room. Properly adjusting the needle valve keeps such venting to a minimum. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 5-11...
The anesthesia ventilator is a volume preset, time cycled, pressure limited ventilator with electronic timing, pneumatic circuitry and independent controls for frequency, inspiratory to expiratory (I:E) ratio, inspiratory flow rate, tidal volume, and inspiratory pressure limit. 5-12 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 77
The ventilator will not function if this pressure drops below 32 psi. Electrical power is supplied by the Narkomed Mobile's AC power source, or, in event of AC power failure, by the backup battery. A fully charged battery can power the ventilator for at least 90 minutes.
The position of the tidal volume indicator can be calibrated for a specific combination of fresh gas flow and equipment compliance by an authorized representative of DrägerService. 5-14 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
15 cmH O. When the control is turned fully clockwise, the peak inspiratory pressure is less than or equal to 120 cmH Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 5-15...
To select air as the drive gas, move the switch to the AIR position. Figure 5-7. Drive Gas Selector Switch DRIVE GAS SELECTOR SWITCH OP00078 VENTILATOR DRIVE GAS FLOWMETER HOUSING 5-16 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Power-On Screen When you turn the SYSTEM POWER switch ON, the Narkomed Mobile performs extensive self-tests on its internal hardware. As these diagnostics are performed, each test and its result appear on the screen. The result, PASS or FAIL, indicates the status of the tested component.
AC / BATTERY FAIL O2 SUPPLY LOW ADVISORY: SERVICE VENT MON PORT A ERROR TID VOL MIN VOL 0.35 RESPIRATORY VOLUME PEAK MONITOR WINDOW MEAN PEEP BREATHING PRESSURE BREATHING PRESSURE TRACE WINDOW MONITOR WINDOW Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 85
(PEEP). • Breathing Pressure Trace Window—Displays a trace, or waveform, of the patient's breathing pressure, and the anesthesia machine's breathing pressure alarm limits (to the left of the waveform). Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
You also enter the Alarm Log by way of the Configure screen. These functions are described in detail later in this section. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Breathing Pressure High Limit Auto OP87008 Configuring You can configure the following parameters on the Narkomed Mobile: • Trace Speed—Speed of the breathing pressure waveform trace: Anesthesia either FAST or SLOW Machine • Alarm Volume—Volume of annunciated alarms • Current Time—The current hour and minute in 24-hour format ( hour:minute ) •...
SELECT Selects a parameter by highlighting it with a box EXIT Exits the Configure screen and returns to the Monitor screen Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Screen passes and no keys are pressed. When you exit the Configure screen, the values displayed on the screen are saved. The monitoring system uses these values until they are changed. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
MINUTE VOLUME LO EXIT Clearing the Alarm Log To delete all of the messages from the Alarm Log, press the CLEAR LOG key. The messages are permanently deleted from the LOG. 6-10 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Note: All continuous audible alarms are automatically silenced for 2 minutes following power-up. During this period, the occurrence of a new alarm produces a non-repeating tone pattern appropriate for that alarm's level of urgency. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-11...
AC / BATTERY FAIL O2 SUPPLY LOW HIGH OXYGEN ADVISORY: SERVICE VENT MON PORT A ERROR CONCENTRATION TID VOL MIN VOL ALARM LIMIT 0.35 LOW OXYGEN PEAK CONCENTRATION ALARM LIMIT MEAN PEEP 6-12 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
3. To save the new value, stop pressing arrow keys until the highlighting box disappears (5 seconds), or press a different alarm Limit key. The new value is saved as the alarm limit. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-13...
Exposure Time 10 seconds > 21% up to 50 seconds 5. When the Narkomed Mobile successfully completes the calibration, pull the inspiratory valve dome plug and reinsert the sensor assembly. 6-14 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Alarm window. If this happens, reconnect the sensor cord to the OXYGEN SENSOR interface beneath the rear panel of the ventilator box and try to calibrate the oxygen sensor again. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-15...
Page 96
If the oxygen sensor is improperly calibrated, it can cause inaccurate measurements. When a calibration gas mixture is excessively rich or lean in oxygen, the Narkomed Mobile will not complete an attempted calibration; however, if the calibration gas is rich or lean but is within certain limits, the Narkomed Mobile will complete the calibration.
(Advisory) • the oxygen sensor enters a noncalibrated state • the Narkomed Mobile is unable to calibrate the oxygen sensor • more than 18 hours have elapsed since the last calibration Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS SERVICE If the Narkomed Mobile detects an internal electronic failure that would VENT MON prevent proper operation, the Advisory message SERVICE VENT MON (Advisory) appears in the Alarm window. If this happens, contact an authorized representative of DrägerService.
A blank MIN VOL display area indicates that a full one-minute history of exhaled volume is not available. • Minute Volume Alarm Limit—Indicates the volume below which an alarm condition occurs. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-19...
While the ventilator is off, these alarms are generated at 30 seconds (Caution) and 60 seconds (Warning). The Narkomed Mobile's volume alarms are automatically enabled when the ventilator power switch is turned to the ON position. A disconnected or damaged sensor causes a sensor failure alarm.
Alarm Messages APNEA- The Narkomed Mobile continuously monitors the expiratory flow in the VOLUME patient breathing system. By processing the expiratory flow pattern, the (Warning/ monitor can determine whether a “valid” breath has occurred. A “valid”...
Page 102
Note: You can disable volume-related alarms by pressing the Breathing Volume Off key. MINUTE Whenever the Narkomed Mobile measures a minute volume less than the VOLUME low minute volume alarm limit, the Caution message MINUTE VOLUME LOW appears in the Alarm window, and an intermittent audible alarm (Caution) sounds.
To the left of the waveform, in small type, are the pressure high and threshold pressure alarm limits. • Breathing Pressure Monitor Window—This small window on the right contains numerical breathing pressure measurements. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-23...
0—20, 0—50, or 0—100 cmH O. If the scale changes, the positions of the waveform and alarm limits also change relative to the new scale. The window below has a scale of 0—20 cmH 6-24 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 105
The following examples illustrate the difference in appearance. (Note that a threshold pressure alarm limit marker is also displayed in these examples.) Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-25...
Limit PRESSURE PRESSURE (cm H O) Breathing Pressure ALARM HIGH ALARM High OFF KEY LIMIT KEY Limit ARROW Auto KEYS THRESHOLD PRESSURE ALARM LIMIT KEY THRESHOLD PRESSURE AUTO SET KEY 6-26 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Warning: The threshold pressure alarm limit should be as close as possible to the sensed peak pressure without exceeding it. For details, see “Threshold Limit Guidelines” and “Sample Threshold Limit Settings” in this section. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-27...
If the alarm limit is on or below the upper limit of the scale, a marker appears next to the alarm limit, and a dotted line extends from the alarm limit horizontally across the scale. 6-28 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 109
Sample Figure 6-18 illustrates the effects of correct and incorrect settings of the Threshold threshold pressure alarm limit. Limit Settings Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-29...
Turning the To turn on the apnea pressure alarm, press the Breathing Pressure On key. Apnea Pressure Alarm On 6-30 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
This alarm condition is cleared when the sensed pressure rises above –10 cmH O. However, the alarm message is extended for 5 seconds to allow the recognition of a momentary subatmospheric pressure condition. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 6-31...
Page 112
Standby, the Advisory message APNEA-PALM STBY (Advisory) appears in the Alarm window. SERVICE If the Narkomed Mobile detects an internal electronic failure that would VENT MON prevent proper operation, the Advisory message SERVICE VENT MON (Advisory) appears in the Alarm window.
Page 113
Disassembling Parts for Cleaning and Disinfection ..7-14 Disassembling the Oxygen Sensor Ass’ly ....7-14 Disassembling the Ventilator Bellows Ass’ly ....7-14 Disassembling the Flow Housing/Transducer Ass’ly ... 7-14 Disassembling the Inspiratory and Expiratory Valves . 7-15 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 114
Cleaning and Disinfecting the Absorber Assembly ........7-22 Cleaning and Disinfecting the Vaporizer Exterior .......... 7-22 Cleaning and Disinfecting the Scavenger Interface ........7-23 Cleaning and Disinfecting the Scavenger Hoses and Bag ......7-23 Reassembly Instructions ................7-24 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Warning: Absorbent is caustic and is a strong irritant to the eyes, skin, and respiratory tract. When replacing the absorbent, take care not to spill its caustic contents. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 116
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Figure 7-1. Disassembled Canister Assembly TOP ABSORBER GASKET CANISTER CANISTER GASKET DUST CUP O-RING DUST CUP OP00094 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
If you do not wait, calibration (the next step) will not be successful. 6. Restore power to the machine and calibrate the oxygen sensor. This procedure is provided in “Using the Monitoring System.” Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
3. Lift the flow sensor off the bracket. 4. Press down on the lever under the flow housing and remove the flow housing/transducer assembly from the electronics housing. 5. Pull both transducers out of the flow housing. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 119
12. Connect the patient hose to the expiratory hose terminal on the flow sensor. 13. Install the connector hose assembly between the flow sensor and the expiratory valve, and secure it by turning the ring nut clockwise. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Note: If your facility requires a disinfection process for the scavenger interface and hoses, remove the scavenger and all 19 mm hoses. Refer back to the cleaning instructions under “Scavenger Maintenance” for disassembly instructions. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
RELIEF VALVE BODY WINGNUT (2X) 5. Disconnect the oxygen sensor connector from the system interface panel. 6. Separate the oxygen sensor housing from the breathing pressure sensor adapter. (It is a press fit.) Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Page 122
7. Disconnect the breathing pressure sensor quick-connect fitting from the system interface panel. 8. Disconnect the quick-connect fitting from the pressure gauge. 9. Separate the pressure sensor adapter from the inspiratory valve dome. (It is a press fit.) 7-10 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 123
10. Turn the nut that connects the pressure gauge to the gauge mount to the absorber assembly counterclockwise. 11. Lift the pressure gauge assembly off the gauge mount. Be sure to keep the o-ring with the pressure gauge assembly. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-11...
Page 124
Take care not to lose the gasket that fits between the expiratory valve and the PEEP valve. 16. Remove the PEEP valve by turning the retaining ring counterclockwise. Take care not to lose the gasket. 7-12 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 125
18. Disconnect the fresh gas hose from the fresh gas outlet. 19. Loosen the wing screw to release the absorber from the absorber pole. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-13...
1. Press down on the lever under the flow housing and pull the flow bling the housing/transducer assembly out of the electronics housing. Flow 2. Pull both transducers out of the flow housing. Housing/ Transducer Ass’ly 7-14 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
(PART OF BREATHING PRESSURE HOSE ASSEMBLY) RING NUT INSPIRATORY VALVE DOME EXPIRATORY VALVE DOME DOME GASKET EXPIRATORY VALVE BODY INSPIRATORY VALVE MOUNT VALVE BODY GASKET PEEP VALVE INSPIRATORY VALVE MOUNT OP00081 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-15...
RETURN TO CD-ROM TABLE OF CONTENTS General Guidelines for Cleaning and Disinfection The frequency, level, and need for disinfection of the Narkomed Mobile is determined by the user facility based on the conditions of use and hospital infection control policy. Draeger Medical recommends using disposable patient breathing circuits that can be replaced after single use.
Page 129
D Wet pasteurization at 70° C for 30 minutes after detergent cleaning E Sodium hypochlorite (5.2% household bleach) 1:500 dilution (100 ppm free chlorine) F Ethyl or isopropyl alcohol (70% to 90%) Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-17...
• Wipe any spills and cleanser off the equipment surfaces immediately. Cleaning\Disi Clean painted, plated, and plastic surfaces of the Narkomed Mobile with a nfecting soft lint-free cloth moistened with mild detergent and water. Follow up with Exterior...
Allow the assembly to drip dry. Ventilator After cleaning, use an EtO process for disinfection. Follow manufacturer's Bellows Bottom guidelines. Assembly Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-19...
After cleaning or disinfecting the inspiratory or expiratory valve assembly, verify that the valve disk and all internal pins are undamaged and that the valve disks are correctly reinstalled. 7-20 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Dust Cup residues for the canister and gasket surfaces. Check the dust cup periodically. Empty and clean it when necessary. Warning: Absorbent is caustic. Avoid contact with the skin and eyes. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-21...
1. Drain the discolored halothane from the vaporizer. 2. Fill the vaporizer with fresh halothane up to the maximum level, then drain completely. 3. Dispose of the drained halothane in accordance with standard practices at your facility. 7-22 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 135
Elbow, and Bag” to select a compatible process. Scavenger Check the hoses frequently for signs of deterioration, including swelling Hoses and tackiness, or cracking. Replace the affected parts when any of these conditions are evident. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-23...
Ensure that the three o-rings on each transducer are not damaged and are properly seated. Slide the flow housing/transducer assembly into the electronics housing. Be sure that it clicks into place. 7-24 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 137
27. Turn the nut that connects the pressure gauge to the gauge mount clockwise. 28. Check the gauge needle’s zero position and adjust it if necessary. 29. Replace the 22 mm hoses and other breathing circuit accessories as illustrated. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-25...
Page 138
Figure 7-12. Replacing 22 mm Hoses and Other Breathing Circuit Accessories (1 of 2) OP00233a VENTILATOR Y-PIECE 22MM MASK CORRUGATED ELBOW HOSE 32 IN. LONG INSPIRATORY VALVE 22MM CORRUGATED HOSE 22 IN. LONG 7-26 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 139
APL valve and ventilator relief valve to the scavenger as illustrated on the following pages. 31. Turn the System Switch on. 32. Perform the Daily Checkout procedure provided in this manual before operating the machine. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-27...
Page 140
Figure 7-14. Replacing the 19 mm Hoses (1 of 2) VENTILATOR RELIEF VALVE 20 IN. 19 MM SCAVENGER HOSE (MAGENTA COUPLING) SCAVENGER APL VALVE MAGENTA BAND EXHAUST SYSTEM 30 IN. 19MM SCAVENGER HOSE (MAGENTA COUPLING) OP00100 7-28 Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 141
RELIEF VALVE 20 IN. 19 MM SCAVENGER HOSE (MAGENTA COUPLING) SCAVENGER OP00082 APL VALVE MAGENTA BAND HOSPITAL SUCTION LINE 3-LITER RESERVOIR BAG (MAGENTA COUPLING) 30 IN. 19MM SCAVENGER HOSE (MAGENTA COUPLING) Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual 7-29...
Page 142
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 143
Absorber System Problem Resolution ....... 8-2 Ventilator Problem Resolution ........... 8-4 Oxygen Monitoring Problem Resolution ......8-5 Respiratory Volume Monitoring Problem Resolution ..8-6 Breathing Pressure Monitoring Problem Resolution ..8-6 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
APL valve resistance is set too Adjust APL valve high. counterclockwise. Automatic ventilation Switch to manual ventilation. malfunction. Ventilator relief valve Contact an authorized adjustment set incorrectly. representative of DrägerService to adjust ventilator relief valve. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 145
Replace gauge. Absorbent is purple. Ethyl violet indicator shows Replace with fresh absorbent. absorbent is exhausted. Other indicators exhibit different color changes. High CO levels. Depleted absorbent. Replace with fresh absorbent. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Absorber manual/ automatic Place selector valve in AUTO inspiration selector valve in BAG position position. Inspiratory flow control setting Increase inspiratory flow control on ventilator too low setting. Frequency too high Decrease frequency. Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Defective sensor housing and Replace housing/cable appears in Alarm window during cable assembly. monitoring Sensor cord is disconnected Insert sensor cord connector into Oxygen Sensor receptacle in the system interface panel. Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...
Hose is blocked or kinked Make sure that assembly is free of obstructions. Condensation accumulation in Drain and reconnect breathing the hose pressure hose assembly. Erratic readings PEEP valve setting Adjust PEEP valve Narkomed Mobile Operator’s Manual Part Number: 4115139-001 Rev: F...
Page 149
RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Spare and Replacement Parts Replacement Parts and Spares Description Part Number Documentation Narkomed Mobile Operator's Instruction Manual....4115139 Absorbent gerSorb (10-liter canister) ..........6750801 ä Absorber System Gasket - Canister Top ...............4105848 Screen - Canister...............1100022 Canister (includes bottom gasket) ...........4105851...
Page 150
O-Ring Set ..........4115147 Narkomed Mobile Operator’s Manual...
Page 151
Cable Translator Adapter ....... . 4114359 Part Number: 4115139-001 Rev: F Narkomed Mobile Operator’s Manual...