GE LOGIQ E10 Basic Service Manual page 616

Hide thumbs Also See for LOGIQ E10:
Table of Contents

Advertisement

D
5750007-1EN, R
IRECTION
MPS removal
Table 8-377 MPS removal
1.
2.
8 - 370
. 1
EV
Steps
NOTE: When the Ultrasound System is in
standby, the MPS (PS) will be warm. This is
normal for the PS.
Disconnect the PS Cables:
P6 - USB to ECB J8.
P8 - 24VDC to Printer, if present.
P9 - 48VDC to XYZ Controller.
Remove the three screws that secure the
PS, using a 5 mm hex wrench.
Ease the PS away from the fang bracket,
located at the top mounting screw, while
supporting the PS at the bottom center.
Pull/slide the entire PS backwards until the
connectors to the Backplane are unseated.
Pull away from the bottom and upward to
remove the PS.
Section 8-12 - MPS (Main Power Supply) replacement
LOGIQ E10 B
S
ASIC
ERVICE
Corresponding Graphic
Main PS and mounting
M
ANUAL

Advertisement

Table of Contents
loading

This manual is also suitable for:

5935000e10

Table of Contents