D
5750007-1EN, R
IRECTION
XY-Mechanism removal
Table 8-225 XY-Mechanism removal
1.
The operator console can drop
unexpectedly when the Z-Mechanism is
released or bumped while working near
this and the Back Cover is off.
Be aware of this if the Z-Mech has
become warn or has weakened.
If you suspect the Z-Mech is warn or has
weakened, it should be replaced.
WARNING
WARNING
Operator console can drop unexpectedly when the mechanism is released.
2.
Lock the Caster Brakes.
3.
Remove the rear Brake Motor Covers, using a T-10 driver.
Carefully disconnect the rear Brake Motors and the ground cables, using a 4 mm hex wrench.
8 - 194
. 1
EV
Step
Section 8-8 - Replacing XYZ Parts
LOGIQ E10 B
ASIC
Corresponding Graphic
Z-Mechanism Manual release lever
S
M
ERVICE
ANUAL