GE LOGIQ E10 Basic Service Manual page 590

Hide thumbs Also See for LOGIQ E10:
Table of Contents

Advertisement

D
5750007-1EN, R
IRECTION
Table 8-355 Light Block (Hard Drive/Network Status) replacement
5.
Re-install all Covers removed.
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section
4-2-3
4-2-7
4-2-7
4-2-4
8 - 344
Section 8-11 - Front End Acquisition / Card Cage parts replacement
. 1
EV
Steps
Table 8-356 Light Block (Hard Drive/Network Status) replacement Functional
Checks
Functional Check
Power ON/Boot Up
Probe/Connectors Checks
B-Mode Checks
Verify Network Activity lights emit through
the Front Cover
Power SHUT DOWN
LOGIQ E10 B
Corresponding Graphic
Debrief Script
LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
S
M
ASIC
ERVICE
ANUAL