General Maintenance Information

Periodic inspections should be performed to ensure safe and
effective operation. It is recommended that a qualified technician
perform a full visual inspection of all components at least every
three months.
The technician should check for the following items:
External damage
Damage to the power supply
Connectors and cable insulation integrity
To ensure patient safety, users should perform a routine
inspection of the GlideScope
use to ensure that all endoscopic components are free of
unintended rough surfaces, sharp edges, protrusions or cracks.
If inspection reveals any faults in the components, contact
Verathon Medical
performed by an authorized Verathon Medical
Caution. Risk of permanent equipment damage.
Do not expose GlideScope
Laryngoscopes or Cobalt Video Batons to
temperatures above 140° F (60°C). Do not
disinfect or sterilize GlideScope
Laryngoscopes or Cobalt Video Batons using
devices such as autoclaves, ultrasonic cleaners,
or pasteurizers. Use of such methods will cause
permanent device damage and void the
warranty. Refer to page 49 for a list of approved
cleaning procedures and products.
User's Manual
®
Video Laryngoscope before every
®
Customer Care. All repairs must be
Cleaning and Maintenance
®
Service Center.
®