Cleaning and Maintenance

Cleaning and Maintenance
Cleaning and Maintenance
Cleaning and Disinfecting the BladderScan
To Clean and disinfect the BladderScan
1. Use a soft cloth dampened with isopropyl alcohol (or an appropriate hospital cleaning
agent) to wipe the Probe until it is thoroughly cleaned.
2. If you use a detergent solution to clean the instrument, remove all residual detergent.
Dry the instrument with a clean, soft cloth.
3. If the instrument needs to be disinfected, dampen a soft cloth in any glutaraldehyde-
based hospital disinfectant solution such as Cidex
Sterilization Products, or Sporocidin
instrument with a dampened cloth.
4. To remove all traces of disinfectant solution and wipe the instrument with a clean soft
cloth dampened in sterile water or cleaning solution. Verathon
the device three separate times to remove all residual disinfectant.
5. Thoroughly dry the instrument with a clean, soft cloth before using.
IMPORTANT! Failure to heed the following warnings may cause device damage not
covered by the BladderScan
Do not immerse the instrument in disinfectant solution.
Do not use CidexPlus
enclosure.
Do not subject any part of the instrument to steam sterilization or ethylene oxide
sterilization.
Changing gel pads between patients is strongly recommended.

Regular Inspections and Maintenance

Verathon Medical
BladderScan
inspection and testing of the instrument to ensure accurate performance in clinical use.
For more information, please contact your authorized BladderScan
local BladderScan
1.800.331.2313 (North America only. International customers, please refer to the contact
information on page 64).
NOTE: ScanPoint
accessing their ScanPoint
Online, please refer to the ScanPoint
page 56
®
BVI 9400 warranty.
®
to disinfect the instrument. CidexPlus will damage the plastic
®
recommends that the BVI 9400 be certified by an authorized
®
Service Center once a year. Certification service includes comprehensive
®
distributor, or Verathon Medical
®
Online customers can maintain device certification via the Internet by
®
account. For more information about using ScanPoint
®
BVI 9400
®
BVI 9400:
®
or Cidex 7
®
from Sporocidin International. Wipe the
®
Customer Care Department at
®
with QuickPrint User's Manual.
®
BladderScan
BVI 9400
®
from Advanced
®
recommends wiping
®
Service Center, your
®
User's Manual