Compliance Level - Paravan WHEELCHAIR Series Operating Instructions Manual

Table of Contents

Advertisement

22.2

Compliance level

The interference immunity levels in accordance with IEC 60601 are fulfilled.
Manufacturer's declaration – electromagnetic interference
The PR 25 is designed to be operated under the electromagnetic conditions specified below. The customer
or the user of the device should ensure that it is used under these conditions.
Emission measurements
HF emissions in accordance with CISPR11 Group 1
HF emissions in accordance with CISPR11 Class B
Emission of harmonics in accordance
with IEC61000-3-2
Emission of voltage fluctuations/flickers
in accordance with IEC61000-3-3
Table 201 in accordance with DIN EN 60601-1-2, 6.8.3.201
150
Compliance
Class A
fulfilled
Electromagnetic conditions –
guidelines
The PR 25 only uses HF energy for its
internal functions. Therefore its HF
emissions are very low, and it is unlikely
that adjacent electronic devices will be
disturbed by these.
The device is designed to be used in all
set-ups including living areas and those
which are directly connected to a public
power grid that also supplies power to
buildings being used for living purposes.
Electromagnetic compatibility (EMC)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pr 25

Table of Contents