After cleaning the different components, put the probe tip and the probe cap into an ultrasonic bath to
guarantee a proper cleaning.
The probe tip should be replaced with a new one when contaminated, completely blocked, or cracked.
If damage occurs to the sealing gasket, the probe must be serviced by an authorized Interacoustics
representative.
D-0114042-D – 2020/03
Sera™ - Additional Information
To clean the larger channel, it is advisable to double the floss to
effectively clean all corners.
To access and clean the larger channel it is necessary to remove the
gasket from inside the probe tip. You can do this using a fine pin. Push
the gasket back into place after cleaning.
DO NOT try to clean the probe tip while it is mounted on the probe.
Only insert the cleaning floss or wire into the base of the probe tip to
ensure wax/debris is pushed out of the probe tip instead of into it. This
also protects the gasket from being damaged. Never use tools to affix
the probe cap!
Remove ear tip from ultrasonic bath after cleaning is completed. The
probe tip should not remain in the bath for longer than 30 minutes.
Reattach the probe tip to the probe and screw the gasket into place.
Do not insert the probe tip into the patient's ear canal without a disposable
ear tip attached.