Teoae Protocols - Interacoustics Sera Additional Information

Hide thumbs Also See for Sera:
Table of Contents

Advertisement

TEOAE protocols

The Sera
device is delivered with one default TEOAE protocol (T00 1.5-4kHz, 3_4, SNR 4 dB). The
TM
protocol on the device can be changed and additional protocols can be added to Sera
HearSIM
software.
TM
The available protocols are:
Protocol name
T00 1.5-4 kHz, 3_4, SNR 4 dB
T01 1.5-4 kHz, 2_4, SNR 6 dB
T02 1.5-4 kHz, 3_4, 80 dB
SPL
D-0114042-D – 2020/03
Sera™ - Additional Information
Parameters
Center frequency bands:
Stimulus type:
Stimulus level:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
Center frequency bands:
Stimulus type:
Stimulus level:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
Center frequency bands:
Stimulus type:
Stimulus intensity:
Stimulus tolerance:
Noise rejection level:
Max test time:
TE detection criteria
Min TE level:
SNR:
Min total OAE:
# bands for a pass:
TM
1.4, 2, 2.8, 4 kHz
Non-linear click
83 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
3 / 4
1.4, 2, 2.8, 4 kHz
Non-linear click
83 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
2 / 4
1.4, 2, 2.8, 4 kHz
Non-linear click
80 dB peSPL
±3 dB
55 dB SPL
60 seconds
-5 dB SPL
4 dB
0 dB
3 / 4
through the
Algorithmic
sensitivity
> 99.6%
> 99.6%
> 99.6%