11.3.1 Cleaning The Touch Screen; Cradle; Preamplifier And Electrode Lead Wires - Interacoustics Sera Additional Information

Hide thumbs Also See for Sera:
Table of Contents

Advertisement

If the Sera
TM
disinfectant wipe
Before cleaning/disinfecting, remove the Sera
Always switch off the device during cleaning
Make sure no liquid enters the connectors of the Sera
plastic
Make sure that the Sera

11.3.1 Cleaning the touch screen

Use a lens cleaning or microfiber cloth to clean the Sera

Cradle

The cradle is not required for performing a screening. To avoid contamination of the cradle and reduce the
need for cleaning/disinfection, we suggest that you do not bring it into the patient care area. If you need to
have the cradle near the patient care area, do not set it on the patient's bedding or hold it in contaminated
hands for testing. Instead, set the cradle on a trolley that is used for transport or on a clean surface near the
patient.
If the cradle becomes contaminated, it can be wiped down with a damp cloth and/or a disinfectant wipe.

Preamplifier and electrode lead wires

The preamplifier and electrode lead wires can be wiped down with a damp cloth or a disinfectant wipe if
necessary.
Be careful that no fluid enters sensitive areas of the preamplifier or electrode lead wire connector.
D-0114042-D – 2020/03
Sera™ - Additional Information
device has become contaminated, it can be wiped down with a damp cloth and/or a
is completely dry before using the device
TM
Before disinfection, remove the Sera
from AC power.
Be careful that no fluid enters sensitive components of the cradle such as the
USB/power connector.
Make sure that the cradle is completely dry before reconnecting the power
cable and reinserting the Sera
The preamplifier and electrode lead wires should be completely dry before
re-use.
from the cradle
TM
or drips into any seams or gaps in the
TM
touchscreen.
TM
from the cradle and unplug the cradle
TM
.
TM
Page 36