Operation
Program canceled
due to a fault
Canceling a pro-
gram manually
Restarting the
program
60

Canceling a program

 If a program is canceled, the items in the machine must be re-
processed again.
 Be careful when opening the door!
The load could be hot. Risk of scalding, burning and chemical
burns.
The program stops prematurely and an error message appears in the
display.
Take appropriate steps to resolve the fault, depending on its cause
(see "Frequently asked questions").
A program which is already running should only be canceled if neces-
sary, e.g. if the wash load is moving in the chamber.
Press and hold the Start/Stop button until the display changes to
the following view:
Cancel program
Use the  and  arrow buttons to select
Pressing the OK button interrupts the program. Entry of a code may
also be required (see "Additional settings/Code").
If no button is pressed for several seconds, or if the process is can-
celed using the  button, the display will revert to the program se-
quence display.
Start the program again or select a new program.
No
Yes