Water inlet and drainage

Problem
Check water intake
Check drainage
148
Possible cause and solution
One or more faucets are turned off.
Turn on the faucets.
There was insufficient water in the machine.
Clean the water intake filters.
Turn on the faucets fully.
Flow pressure at the water connection is less than 4.35 psi
(0.3 bar/30 kPa).
Contact a qualified plumber.
A program was canceled because the water in the wash
chamber is only being pumped away slowly or not at all.
- The drain hose is blocked.
Remove any kinks or large loops in the drain hose.
Start the program again.
- The filters in the wash chamber are blocked.
Clean the filters in the wash chamber.
 Danger of injury from glass shards, needles etc.
which are retained in the filter.
Start the program again.
- The drain pump or non-return valve is blocked.
Clean the supply line to the drain pump and the non-re-
turn valve.
Start the program again.
- The drainage system cannot accommodate the water be-
cause it is blocked.
Contact a qualified plumber.