Retrofitting The Large-Surface Filter - Miele PG 8583 CD Operating Instructions Manual

Laboratory glassware washer for laboratory glassware and utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Retrofitting the
large-surface fil-
ter
DI water connec-
tion for a pressur-
ized system with
4.4 - 145 psi
(30-1,000 kPa)
gage pressure
DI water connec-
tion for a non-
pressurized sys-
tem below 4.4 psi
(30 kPa) gage
pressure
DI water ring line
 Do NOT shorten or otherwise damage the inlet hoses (see dia-
gram).
Also refer to the supplied installation diagram!
If the water contains a high level of insoluble components, a large-
surface filter can be installed between the faucet and the water inlet
hose.
The large-surface filter is available from Miele Customer Service.
This machine is suitable for a pressurized DI system operating be-
tween 4.4 - 145 psi (30-1,000 kPa). If the water pressure is below 29
psi (200 kPa) the water intake duration will be automatically in-
creased.
The pressure tested hose for DI water, coded green, has a ¾"
threaded union for connection to the onsite faucet for DI water.
 If the machine is not going to be connected to DI water, the DI
water supply has to be deactivated by Miele Service. The intake
hose remains on the back of the machine.
DI water connections below 4.4 psi (30 kPa) gage pressure require
the installation of an external booster pump, which can be requested
through Miele Service. Installation of the pump must be carried out by
Miele Service.
The machine can be connected to a ring line system for DI water. For
this purpose, it must be technically adapted and the controls reset by
Miele Service.
Please contact Miele Service for further information.
Plumbing
161

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

62858350usaPg 8583

Table of Contents