User Manual
A3000
AC Power Source
Example :[SOURce:]PHASe:ON?
Parameter :<NR2>, effective range:0.0 ~ 359.9
Parameter returns
[SOURce:]PHASe:OFF
Function :Set (query) waveform stop phase angle set at main manual.
Example :[SOURce:]PHASe:OFF?
Parameter :<NR2>,effective range:0 ~ 360.0, 360.0 representing turn off immediately
Parameter returns
[SOURce:]LIST:COUNt
Function : Set (query) how many times to execute effective sequences in LIST mode.
Example : [SOURce:]LIST:COUNt?
parameter : <NR1>,effective range: 0 ~ 10000
Parameter returns
[SOURce:]LIST:DWELl
Function : Set (query) execution time of 10 individual sequences in LIST mode.
Example : [SOURce:]LIST:DWELl?
parameter :<NR2>, ..., <NR2> effective range:0 ~ 60000 (unit: msec)
Parameter returns
[SOURce:]LIST:SHAPe
Function : Set (query) waveform buffer used for 10 individual sequences in LIST mode.
Example : [SOURce:]LIST:SHAPe?
67
│ www.deltaww.com
[SOURce:]PHASe:ON 200.5
:<NR2>
[SOURce:]PHASe:OFF 250.5
:<NR2>
[SOURce:]LIST:COUNt 100
:<NR1>
[SOURce:]LIST:DWELl 60000 1 23 95
:<NR2>, ..., <NR2>
[SOURce:]LIST:SHAPe A B A A A