3.5.3 Voltage Level Setting
A3000 SERIES provide two output voltage levels used for low voltage and high voltage
applications. At 300V level, AC source can provide 16.0A in total;At 150V level 32.0A in
total can be provided, maximum output power is 3000VA.
When switched from 300V to 150V level, AC source will check all voltage settings, and
change the setting to 150V if any voltage setting checked is higher than 150V.
When switched from 150V to 300V level, AC source will check all I Limit settings, and
change the setting to 16.0A if any current setting checked is higher than 16.0A.
Following processes demonstrate how to change output voltage level (300V → 150V):
1. Press
2. Rotate knob to change <300V> to<150V>, then press
3.5.4 Voltage Sense
Voltage Sense determine output voltage measurement point is at AC output terminal
(Vout) or input terminal at product under testing (Vsense). If it is set at Vsense, Remote
Sense wire must be connected first, otherwise compensate error will occur. AC source will
compensate the voltage drop of the output cable if Vsense is selected.
Following processes demonstrate how to set Volage Sense as Remote Sense:
Save
1. Press
22
│ www.deltaww.com
Save
Save
Recall
Recall
or
to move cursor to Voltage Level
▲
▲
▼
▼
Save
Recall
Recall
or
to move cursor to Voltage Sense
▲
▲
▼
▼
to confirm setting
Enter