Dc Power Cable - GE Druck ADTS 405 R Mk2 Instruction Manual

Air data test set
Table of Contents

Advertisement

Chapter 5. Maintenance
f.
Measure the insulation resistance between Live and Neutral.
g.
In all cases the insulation resistance must exceed 2 MΩ.
ADTS Pin
1
4
EARTH

5.5.3 DC Power Cable

Note: For part AS405-31-1891M0.
a.
Prepare to measure continuity using the DMM (item 1, Table 5-3) set to the applicable
resistance range.
b.
Measure the continuity between corresponding ADTS connector pin and the
conductors at the open end of the cable using the connectivity Table 5-5 below. The
measured resistance should not exceed 0.5 Ω.
Note: This resistance value is the expected maximum value for the 5 m (16 ft)
cable length supplied.
c.
Prepare to measure insulation using the Insulation tester (item 2, Table 5-3), set to
the 500 V range. This test will be most practically performed at the open end of the
cable.
d.
Measure the insulation resistance between Gray (+28 V) and Green (Protective
Ground).
e.
Measure the insulation resistance between White (0 V) and Green (Protective
Ground).
f.
Measure the insulation resistance between Gray (+28 V) and White (0 V).
g.
In all cases the insulation resistance must exceed 2 MΩ.
ADTS Pin
EARTH
56 | ADTS 405 R/F Mk2 Instruction Manual-English
Table 5-4: AC Input Connector
European Color
Brown
Blue
Green/Yellow
Table 5-5: DC Input Connector
Cable Color
1
2
USA Color
Black
White
Green
Gray
White
Green
© 2017 General Electric Company – All Rights Reserved.
Function
Live
Neutral
Protective Ground
Function
+28 V
0 V Return
Protective Ground