Chapter 5. Maintenance
f.
Measure the insulation resistance between Live and Neutral.
g.
In all cases the insulation resistance must exceed 2 MΩ.
ADTS Pin
1
4
EARTH
5.5.3 DC Power Cable
Note: For part AS405-31-1891M0.
a.
Prepare to measure continuity using the DMM (item 1, Table 5-3) set to the applicable
resistance range.
b.
Measure the continuity between corresponding ADTS connector pin and the
conductors at the open end of the cable using the connectivity Table 5-5 below. The
measured resistance should not exceed 0.5 Ω.
Note: This resistance value is the expected maximum value for the 5 m (16 ft)
cable length supplied.
c.
Prepare to measure insulation using the Insulation tester (item 2, Table 5-3), set to
the 500 V range. This test will be most practically performed at the open end of the
cable.
d.
Measure the insulation resistance between Gray (+28 V) and Green (Protective
Ground).
e.
Measure the insulation resistance between White (0 V) and Green (Protective
Ground).
f.
Measure the insulation resistance between Gray (+28 V) and White (0 V).
g.
In all cases the insulation resistance must exceed 2 MΩ.
ADTS Pin
EARTH
56 | ADTS 405 R/F Mk2 Instruction Manual-English
Table 5-4: AC Input Connector
European Color
Brown
Blue
Green/Yellow
Table 5-5: DC Input Connector
Cable Color
1
2
USA Color
Black
White
Green
Gray
White
Green
© 2017 General Electric Company – All Rights Reserved.
Function
Live
Neutral
Protective Ground
Function
+28 V
0 V Return
Protective Ground