7.3.3 SETUP, [LIMITS]

The ADTS 405 Mk2 is preprogrammed with the manufacturer's STANDARD, CIVIL and MAX
limit sets. Different sets of additional aircraft limits can be programmed using the CONFIG
mode. SETUP, [LIMITS] allows the selection of the set of limits to be used, each set of limits is
identified by its aircraft name.
Note: The ADTS 405 Mk2, when delivered, contains "Standard" and "Civil" limits; to set
and store "Max" limits, for the first time, enter configuration. The "Max" limits depend on
the range of the pitot channel either 850 or 1000 kts, see Section 1.3, "Operating Limits,"
on page 2.
1.
Use [NEXT] or [PREV] to step through the aircraft names select [SEL] and press QUIT
when the display shows the required name.
2.
Use CONFIG, [LIMITS] to program the sets of limits. Each set of limits includes:
• Max ALT
• Min ALT
• Max CAS
• Min CAS
• Max Mach
• Max ROC
• Max Rate CAS
• Max Ps
• Min Ps
• Max Qc
• Min Qc
• Max Rate Ps
• Max Rate Qc
• ARINC limits on or off
© 2017 General Electric Company – All Rights Reserved.
O 4°C
O 20°C
O 60°F
2
ADTS 405 R/F Mk2 Instruction Manual-English | 83